. Acuson Sequoia C512. Acuson Sequoia C256. Sonosite Titan. Philips IE-33. SONOS 4500/5500/7500. Acuson Aspen. GE Vivid Systems. and any other ultrasound device with 2D cardiac transducers.
![]()
Re: Problem with Acuson Sequoia 512June 23, 2011 06:28Dinu, hi! You can try: press MULTIHERTZ and V KEY during power on = you may have more informations about boot sequence untill loading stops. The most likely causes of a Sequoia system hang: 1. SSP failure on the RDP board 2. Missing Clock signals from RX. (Secondary clocks from CN) 3.
Faulty Hard Drive 4. Misconfiguration 5 Corrupt data 6. SCSI problem 7.
RI or AUX Preamp not seated correctly I hope this will help you. Do not replace the following combinations of FRUs at the same time: 1. Hard drive and RDP board 2. Hard drive and PIC board II. Before replacing an FRU, it is recommended that the system be in the service enabled mode. Re: re: re: re: re: Re: Problem with Acuson Sequoia 512August 25, 2016 04:36We are getting the system error that the MO Disk was not ready after just booting the system up while it still has the boot up gray screen with picture of the tree and will not go past this system error into the imaging screen, however it was not doing this earlier today and just started doing this. My colleague thinks he hit a button on the underside of the keyboard earlier that reset the system.
Since then we have tried to start it up and first got “System Error MO Disk Not Ready.” But now when we boot up the system with the MO Disk in the drive it now shows the system error that “lynx.os.ssp cannot be found.” I was hoping that maybe you guys might have some ideas about what was going on and how we could address this issue to fix it. Thank you in advance for your help. Please reply here or email us at [email protected]. Re: Re: Problem with Acuson Sequoia 512April 07, 2016 03:14Hi Marko I have an Acuson Sequoia C512. When I switch on it, on the screen is displayed the message 'System starting.'
, but it isn't start ever. All the buttons are 'frozen' and to switch off must pull on the plug. When I put in the new HDD with the same characteristics as the old one, on the screen is displayed massage:ERROR 30'. I can't reload SW because as I mention all the butons are 'frozen' Is SW corrupted and a new HDD must be loaded? Has it problem with PIC or RDP or both? May you tell me what happened?